ru
en
F4 room, devices, etc.
2019-2024
Shielded room (11.2019)
New tables and device rack
mounting Tenma power supplies
device rack: grounding bar, power supplies
device rack: Pico oscilloscopes and ADC
first run of the cryostat
13.12.2019, installing computers
connector box for cryostat heaters
Oxford He level meter
I like this wiring
PicoLog ADC24
PicoScope with new grounding
PicoScope with new grounding
Shelves in the shielded room (09.2020)
soldering holders for shelves
flux and grade D solder (Sn/Pb=70/30)
shelves
Pump room (10-11.2020)
1/16" pipe adapters made in the workshop
new Swagelock connectors for N2 traps
exhaust for air pumps
N2 circulation traps
new copper pipes to the traps
pump room
old valves for cooling water
Raspberry Pi for pressure measurements (03.2021)
Installing generators at the back side of the rack (03.2021)
Keysight mounting kit,
[dxf]
Gas handling system (03.2021)
bending 6mm pipe
KF to 6mm Swagelock adapters welded in the workshop
new panel for circulation pumps
new location for circulation pumps (roots moved up, rotary under it)
H2 filter installed near circulation traps
N2 trap for 3He sample
cryostat
cryostat flange, thermometer/heater connector box
LC filters for magnets (03.2021)
capacitors
filter box in the cryostat
Femto lock-ins (03.2021)
A small 19" upper rack installed on the cryostat flange. Two blocks
(5 lock-ins each) installed, original power supplies removed, output connectors
(10 pin Lemo) and power connectors (3 pin XLR) added on back panels.
upper rack with one lock-in block installed
output connector
power wires
PCB and power connector
power extensions
second block installed
Lock-in output is measured with PicoLog ADC24
adapter for ADC
soldering dividers (1:4, 10->2.5V)
lock-in output cables (10 pairs)
ADC in the rack
Drive boxes (04.2021)
Drive box contains input BNC connector, separation transformer 6:1 for
separating ground, 100R calibration resistor with BNC connector, current
source (0, 1k, 10k, 100k, 1M, inf resistors with switch), 2 output BNC
connectors.
Part list
- RS 5173951 RS PRO Black Die Cast Aluminium Enclosure, Shielded, 114.5 x 63.6 x 30.3mm 8.6 10
- RS 2106425 Through Hole Audio Transformer 600 100mW Hz @ 300, 1mW Hz @ 30 12.86 10
- RS 2106469 Transformer Screening Can 5.31 10
- RS 8779874 Grayhill Rotary Switch 12.43 10
- RS 8780240 Grayhill Rotary Switch Knob for use with Encoders, Rotary Switch 2.27 10
- RS 7548770 TE Connectivity 1M Metal Film Resistor 0.25W 0.1% YR1B1M0CC 0.449 15
- RS 8307646 Vishay 100k Metal Film Resistor 0.125W 0.1% PTF56100K00BZEK 1.074 12
- RS 8307422 Vishay 10k Metal Film Resistor 0.125W 0.1% PTF5610K000BYEK 1.223 30
- RS 8306779 Vishay 1k Metal Film Resistor 0.25W 0.1% PTF651K0000BYEK 1.496 12
- RS 8498953 Vishay 100 Metal Film Resistor 0.125W 0.1% PTF56100R00BYEK 1.223 12
- RS 5464910 RS PRO 50 Straight Panel Mount BNC ConnectorBulkhead Fitting, jack 2.526 30
- RS 2843792 RS PRO Male BNC to Male BNC RG58 Coaxial Cable, 50 9.52 15
one of original boxes
new box
new box with fixed 100k resistor
drive boxes in the rack
second ADC unit installed
Top rack (12.2021)
Fixing Femto lock-ins (11-12.2021)
The problem was a huge cross-talk between reference and input.
It was even possible to drive wire through the input and measure
resonance with fully disconnected drive cables.
The most efficient update was putting aluminium shield on every
lock-in module. Also I made new back PCB panels and installed coaxial
cables with SMA connectors for input and reference lines between PCB and
front panel.
Back PCB:
KiCAD project
Gerber files for ordering
lock-in module
old front panel connection: unexpected connections between lines possible
old back PCB
front panels
metal shield added
new back PCB
modified front panels: BNC feedthroughs for IN, REF, MON; PCB headers for output
front panels connected
output connector
power connector
Misc (07.2022)
multiplexor board
bath heater box ("green connector")
relay for remote control added
Adding five more lock-ins, tinyXLR connectors (06-07.2022)
15 generators
device rack
BNC to tiny XLR adapters for lock-in inputs
soldering tinyXLR to shielded twisted pair
drive cable (15 BNC)
drive and reference cables in the rack
all outputs connected to ADC
15 drive boxes
drive box with added tinyXLR output
new drive box (no BNC output)
drive box with switch
original cryostat connectors removed
upper rack
main rack
Final connections of vibrating wires (09.2022)
Demag power supply
Demag power supply: Cryogenic PS120A.
2019.11.29 - installed in the rack
2023.01.18 - rate control resistor replaced
2023.01 - an additional connector added for parallel connection
of an additional power supply before the thermally stabilize shunt resistor.
This was really useful for remote current control.
front view
top view
replacing rate control resistor
Misc (03.2023)
replacing batteries in UPS
new panel for magnet wires